Characteristics
Bandwidth - DC to 50 MHz.
Max DC + Peak AC Current - 15 A.
Max Peak Pulse Current - 50 A.
Minimum Sensitivity - 10 mA/div.
DC Accuracy -
±1% (with probe calibrator).
±3% (without probe calibrator).
Max Bare Wire Voltage - 300 V (CAT I).
Typical Characteristics
Insertion Impedance -
0.002 Ω at 1 kHz.
0.1 Ω at 5 MHz.
Propagation Delay - 17 ns.
Max Conductor Diameter - 3.8 mm (0.15 in.).
Electrostatic Immunity - (per IEC 801-2) >8 kV.
Ampere-sec Product - 500 × 10-6 Amp×Sec.
Rise Time - ≤7.0 ns.
Physical Characteristics
Weight - 0.19 kg (0.44 lb.).
Cable Length - 2.2 m (7.2 ft.).
Power Requirements - TCP202 is powered by the TDS oscilloscope through the TEKPROBE interface.
Safety
UL 3111-1, CSA1010.1, EN61010-1, IEC61010-1, EN61010-2-032, IEC61010-2-032.
Recommended Oscilloscopes
TDS520B; TDS520C; TDS520D; TDS540B; TDS540C; TDS540D; TDS580C; TDS580D; TDS654C, TDS680C, TDS684C, TDS694C, TDS724A; TDS724C; TDS724D; TDS744A (S/N B040100 or higher); TDS754A; TDS754C; TDS754D; TDS784A (S/N B040100 or higher); TDS784C; TDS784D; TDS794D; TDS3000 Series; TDS3000B Series, TDS5000 Series; TDS5000B Series; TDS6000 Series and TDS6000B Series with TCA-BNC adapter; TDS7000 Series and TDS7000B Series with TCA-BNC adapter.
For TCP202 operation with other TDS400, TDS500, TDS600, and TDS700 Series oscilloscopes, the scale factor will be correct but measurement will be indicated as V instead of A.
Increase the accuracy of your current measurements
The accuracy of the TCP202 AC/DC current probe is normally 3%. However, the accuracy may be improved to 1% when the probe is calibrated on certain TDS oscilloscope models using the optional Calibrator P/N 015-0672-50. This improved accuracy is available on the TDS Oscilloscopes Series TDS500, TDS600, and TDS700, however the optional calibrator is only functional with the TDS500, TDS600, or TDS700 Series oscilloscope calibrator pins.
Ordering Information
TCP202
DC Coupled Current Probe.
Includes: Instruction manual (070-9542-xx), reference manual (070-9543-xx).
Recommended Accessories
Probe Calibrator - (015-0672-50) Use with TDS510A, TDS520B, TDS520C, TDS520D, TDS540B, TDS540C, TDS540D, TDS580D, TDS654C, TDS680C, TDS684C, TDS694C, TDS724A, TDS724C, TDS724D, TDS754A, TDS754C, TDS754D, TDS784A, TDS784C, TDS784D, TDS794D for improved accuracy.
Related Products
1103 - TEKPROBE® Power Supply (for scopes without TEKPROBE BNC interfaces).
ADA400A - Differential Preamplifier.
P5205 - High-voltage Differential Probe.